ASTIGMATISM MANAGEMENT


New technologies allied to improved surgical techniques are helping surgeons to deliver more predictable, accurate and reproducible outcomes in the treatment of astigmatism according to a number of studies presented here.
“For better outcomes with toric IOL implantation we know that we need more accurate surgery and the latest generation of technologies will certainly help us to achieve that,†Roger F Steinert MD told delegates attending the XXIX Congress of the ESCRS.
Dr Steinert, the Irving H Leopold professor and chair of ophthalmology, professor of biomedical engineering and director of the Gavin Herbert Eye Institute at the University of California, noted the critical importance of accurate toric IOL alignment in treating astigmatic patients.
“If the alignment is off by just 15 degrees, we lose half of the IOL effect and 30 per cent misalignment means that we have lost all of the effect as well as picking up some higher order aberrations,†he said.
The traditional method of marking the eye for toric IOL implantation introduces a number of variables that undermine the likelihood of obtaining precise and reproducible outcomes, said Dr Steinert.
“Marking the limbus must be performed preoperatively with the patient in an upright position. Once the patient is draped, it can be hard to determine the coronal or horizontal plane. Cyclotorsion is another big problem and it is the rule, not the exception, in a supine patient. We did a study a few years ago and found that the mean deviation was four degrees, but that it was not unusual to see as much as 15 degrees of cyclotorsion, so patients will vary quite a bit,†he said.
For the marking of the eye, Dr Steinert said that there are a wide variety of instruments currently on the market such as the Mendez gauge, the Dell marker or the Steinert-Rumison marker, which do the job adequately. He advised surgeons using ink marks to select the 3 o’ clock and 9 o' clock positions rather than 6 and 12 o’ clock, because the latter positions require the surgeon to manipulate eyelids and may lead to greater inaccuracy.
Taken collectively, all of these factors lead to an increased chance of eventual toric IOL misalignment.
“The variables start to add up with the estimation of what is truly horizontal and vertical, the possible spread of the ink marks and problems of cyclotorsion. This is really what is driving the technological drive to see if we can come up with something better,†he said.
New devices
Dr Steinert said that new technology platforms such as the Osher Toric Alignment System (Haag-Streit Inc), the TrueVision 3D Visualization and Guidance System, Callisto (Carl Zeiss Meditec AG) and the ORA(WaveTec Vision) should all help to significantly reduce the risk of misalignment.
“The technology now coming onto the market is showing us some pathways for improvement but we need to continue to balance clinically significant improvements against cost, as there is capital outlay, user fee and time involved in these new devices as well,†he concluded.
The need to move from subjective refractive assessment to more objective measurements was also emphasised by Roberto Zaldivar MD in a separate presentation.
Dr Zaldivar focused in particular on the benefits of using the OQAS (Optical Quality Analysis System, Visiometrics S.L, Spain), noting that the device offers an effective means to measure accurately and objectively patients’ quality of vision.
“This device introduces a new concept that we call super refraction. Most ophthalmologists consider normal vision as 20/20 but we know now that all 20/20 visual acuities are not the same. When an optical system is subject to diffraction plus aberrations plus scattering, all of these affect optical quality. The OQAS helps us to explain why some 20/20 patients see better than others,†he said.
Based on research by Pablo Artal PhD and Jaume Pujol PhD, the OQAS system consists of laser diagnostic sensory equipment, a computer workstation and custom-designed software. The device uses a double-pass technique to analyse the size and shape of a punctual light source imaged on the retina after passing twice through the ocular media.
The analysis of the light provides users with the point spread function (PSF), optical scattering index (OSI), the modulation transfer function (MTF) of the eye, as well as tear film quality.
The PSF, OSI, and MTF and tear film analysis are highly accurate measures of image quality, which provide an objective assessment of the presence of visual aberrations and light scattering, said Dr Zaldivar.
“The OQAS allows for objective measurement and has a clear advantage over traditional aberrometers in that it can measure higher order aberrations and scatter. Using this approach, we obtain better results in post-LASIK eyes, irregular corneas, corneas with rings, and normal eyes as well,†he said.
Summing up, Dr Zaldivar said that the OQAS devices gives surgeons an exciting tool for improving refractive and surgical results.
“It is mandatory to keep improving our preoperative information in order to obtain better visual quality for our patients postoperatively. With this system, I think we are finally moving to an exciting new era of objective refraction,†he concluded.
Latest Articles
Simulators Benefit Surgeons and Patients
Helping young surgeons build confidence and expertise.
How Many Surgeries Equal Surgical Proficiency?
Internet, labs, simulators, and assisting surgery all contribute.
Improving Clinical Management for nAMD and DME
Global survey data identify barriers and opportunities.
Are Postoperative Topical Antibiotic Drops Still Needed?
Cataract surgeons debate the benefits of intracameral cefuroxime prophylaxis.
Emerging Technology for Detecting Subclinical Keratoconus
Brillouin microscopy shows promise in clinical studies.
Knowing Iris Repair: Modified Trifold Technique
Part eight of our series covers the modified trifold technique for large iris defects.
It’s All About Biomechanics!
Increasing the pool of patients eligible for refractive surgery.
Uncovering More Safe and Quick Options
Different strategies, such as PresbyLASIK, can offer presbyopes good outcomes.
Topography-Guided PRK for Keratoconus
Improving visual acuity in patients with keratoconus.
Defining AMD Treatment Protocol
Treatments trending to fewer injections for better results.