ESCRS - PO074 - ASSESSING VISUAL OUTCOMES OF A NOVEL TRIFOCAL INTRAOCULAR LENS DESIGN USING A CLINICAL ADAPTIVE OPTICS SIMULATOR

ASSESSING VISUAL OUTCOMES OF A NOVEL TRIFOCAL INTRAOCULAR LENS DESIGN USING A CLINICAL ADAPTIVE OPTICS SIMULATOR

Published 2025 - 29th ESCRS Winter Meeting

Reference: PO074 | Type: Presented Poster & Poster | DOI: 10.82333/km6w-1b70

Authors: Walter D. Furlan* 1 , Anabel Martinez-Espert 1 , Salvador García Delpech 2 , Diego Montagud Martínez 3 , Vicente Ferrando 4 , Juan A. Monsoriu 4

1Departamento de Óptica,Universitat de Valencia,Burjassot,Spain, 2Clínica Aiken,Fundación Aiken,Valencia,Spain, 3Departamento de Óptica,Universitat de Valencia,Burjassot,Spain, 4Centro de Tecnologías Físicas,Universidad Politécnica de of Valencia,Valencia,Spain

Purpose

To evaluate in real patients the visual performance of new MIOL design using an adaptive optics visual simulator. The MIOL is a trifocal-EDoF design developed to enhance intermediate visual performance. 

Setting

Clínica Aiken, Diffractive Optics Group, University of Valencia (Spain)

Methods

The Visual Adaptive Optics Simulator (VAO, Voptica SL, Murcia, Spain) was employed to obtain the visual acuity defocus curve from 25 eyes of healthy volunteers virtually implanted with the Devil's MIOL. Additionally, images of optotypes were obtained with a digital camera to qualitatively illustrate patients' perception at different vergences.

Results

Mean visual acuity values were better than 0.18 logMAR across the entire range of vergences (-0.5 D to +3.5 D), resulting in 0.03 logMAR at 0.00 D vergence, 0.10 logMAR at +1.50 D vergence, and 0.11 logMAR at +3.00 D vergence. The experimental outcomes compared favorably with numerical results obtained previously in a model eye.

Conclusion

The new MIOL design provided very good visual acuity outcomes in actual patients, as assessed with the Visual Adaptive Optics Simulator. Prototypes of this MIOL are currently under development.