Precision Pulse Capsulotomy To Reduce The Risk Of Komplikations Of Advanced Cataracts
Published 2022
- 40th Congress of the ESCRS
Reference: PO103
| Type: Free paper
| DOI:
10.82333/wkm7-k180
Authors:
Magda Beatrice Rau* 1
1Privatklinik Dr.Rau,Augenklinik Cham,Cham,Germany
Purpose
To evaluate the safety and efficacy of precise capsulotomy device Zepto (Mynosys Cellular Devices) in advanced cataracts.
Setting
Augenklinik Cham,Germany
Methods
From January 2021- January 2022, 26 eyes of 26 patients with advanced cataract were included in prospective study.The inclusion criteria were advanced posterior polar cataract,mature,white and hard nuclei cataract (cataracta brunescens)Included were 6 eyes with advanced cataract with narrow pupil
Excluded wereeys with pseudoexfoliation,insufficiens of zonula fibers and lens dislocation.
Results
Precision Pulse Capsulotomy created in all cases a circular anterior capsulotomy of a precise diameter.In one case anterior capsula tear occured,even so the intracapsular stable implantation of IOL was possible.The tear was caused with sharp nucleus fragment.In no case posterior capsular rupture occured.BSS floats the Zepto tip off the capsule enter the subcapsular space in a 360 degree fluid wave that begins at the capsulotomy egde and separates LECs from the anterior capsule.This facilizate the phacoemulsification and the removal,aspiration of epinucleus and cortex and accelerates the whole procedure In 6 cases with narow pupil the iris extension device was used after the CR.
The achieved position of all IOL in capsular bag was stable.
Conclusions
Precision pulse capsulotomy is a safe effective procedure for the advanced cataract, Zepto reduces the risk of capsula ruptur and facilizes removal
of the cortex and epinucleus .
It is as well convenient for cases with small pupil