Ocular Measurement Comparison Of Swept-Source Optical Coherence Tomographer Argos, Scheimpflug Imager Pentacam, And Partial Coherence Interferometer Iolmaster
Published 2022 - 40th Congress of the ESCRS
Reference: FPS03.06 | Type: Free paper | DOI: 10.82333/pz2s-jx26
Authors: Antti Viljanen* 1
1Silmaasema,Turku,Finland
Purpose
To evaluate agreement between various biometric parameters using three measurement devices.
Setting
Silmaasema Turku, Finland
Methods
Results
Correlations of keratometry was 0,976-0,997, p<0,001 for all. K1 difference between ARGOS and Pentacam was -0,07D/-0,10D (right/left), p<0,04/p<0,04. K2 difference was -0,07D/-0,05D, p=0,14/p=0,28. Correlations of CYL, J0 and J45 was between 0,625-0,928, p<0,001 for all. ARGOS CYL medians were 0,75D/0,84D, and Pentacam 0,70D/0,80D, p=0,46/p=0,07. ARGOS J0 medians were 0,17/0,21, and Pentacam 0,19/0,19, p=0,18/p=0,92. ARGOS J45 medians were -0,12/-0,03, and Pentacam -0,13/0,01, p=0,83/p<0,001. Correlations of ACD was between 0,935-0,997, p<0,001 for all. ARGOS ACD was 3,25mm/3,27mm, and Pentacam 3,15mm/3,18mm, p<0,001/p<0,001. ARGOS AL medians were 23,51mmm/23,52mm, and IOLMaster 23,47mm/23,49mm, p=0,10/p=0,06.